Müller & Sebastiani Elektronik GmbH
Deutsche Version
Sitemap
 
  Test & Measuring Devices -> Zeus Defibrillation Simulator
  Start  | Product Information  | Accessories  | Pictures  | Specification  | Calibration  | Instruction Manual  | FAQ  | Online Shop

Zeus Defibrillation Simulator: Specification





Enclosure material Solid plastic, front panel made of aluminium
Dimensions
(D x W x H)
Approx. 24 x 27 x 24 cm
Weight Approx. 7.0 kg
LCD display 3.5 digit liquid-crystal display
Electronics Supply voltage: 230 V / 250 mA
Test voltage can be increased steplessly up to 5 kV
Automatic discharge at 5.2 kV
Experimental setup in accordance to EN 60601-1
Integrated 10 Hz sine-wave generator
Integrated measuring circuit for the connection of an oscilloscope and a test clip (dynamic testing)
Internal discharge for interruption of the supply voltage and test interruption
Maximum number of capacitor discharges: 10.000
Maximum discharge frequency of the capacitor: 2 discharges/min
Discharge circuit Zeus: L = 500 uH, R = 50 ohm
Zeus 25/400: L = 25 mH, R = 400 ohm
Surroundings Temperature: 10 – 40 ° C
Air humidity: 10 – 75%, noncondensing
Atmospheric pressure: 75 kPa – 106 kPa

 

© 2004 - 2017 Müller & Sebastiani Elektronik GmbH - Webmaster | Imprint