Test & Measuring Devices -> phantom 320 ECG Simulator |
Start
| Product Information
| Pictures
| Specification
| Calibration
| Brochure
| Instruction Manual
| FAQ
| Online Shop
phantom 320 ECG Simulator: Specification
|





|
| 12 leads |
I,II,III,avR,avL,avF,V1,V2,V3,V4,V5,V6 |
| 14 connections |
4 mm banana plugs and pushbuttons
Identification and colour code:
DIN EN 60601-2-51 |
| Digital storage |
1,6 Mbit |
| Amplitude resolution |
8 Bit |
| Scanning frequency |
256 Hz |
| Signal bandwidth |
0 – 120 Hz |
Signal amplitudes (min - max) |
Lead I + 1.35 – 1.49 mV
Lead II + 2.35 – 2.49 mV
Lead III + 0.95 – 1.06 mV
Lead V1 - 2.12 – 2.30 mV
Lead V2 - 0.55 – 0.61 mV
Lead V3 + 1.03 – 1.14 mV
Lead V4 + 2.11 – 2.21 mV
Lead V5 + 1.75 – 1.85 mV
Lead V6 + 1.40 – 1.51 mV |
| Time base |
Quarz 32768 Hz |
| Output impedance |
Electrode connections: 20 Ohm
1V outputs: 100 Ohm |
| Waveforms |
8 sinus rhythms (normal QRS)
30,45,60,75,90,120,150,180 BPM
7 supraventricular arrhythmia
9 ventricular arrhythmia
3 pacemaker
2 ST segments (elevation, depression)
Each with horizontal, ascending and
descending ST wave
Bradycardia
Tachycardia
ECG with artefacts
ECG with interference 50/60 Hz |
| Signal width |
PQ, QRS, QT depending on RR interval
(after Lepeschkin) |
| Battery |
Standard 2 x 1.5-V LR 6 |
| Battery life |
Approx. 100 hours continuous operation |
| Surroundings |
Working temperature 0 bis 40 °C
Storage temperature -20 bis 80 °C
Air humidity: 10% - 90% noncondensing
Barometric pressure 700 - 1500 hPa |
| Dimensions (LxWxH) |
150 x 100 x 40 mm |
| Weight |
400 grams, including battery |
| | |