Medical Devices -> MS300/310 Polygraph |
Start
| Product Information
| Pictures
| Software
| Specification
| Brochure
| Instruction Manual | Quick Reference
|
Online Help
MS300/310 Polygraph: Specification
|




|
Measurements
|
Signal/Parameter |
Sampling Rate |
Basic Diagnostics (MS300/310) |
Pulse Rate |
1 per second |
SpO2 |
1 per second |
Signal Quality (Pulse) |
1 per second |
Pulsation Strength (Total Pulse Modulation) |
1 per second |
Pulse wave (Plethysmogram) |
50 per second |
Respiration (Thermistor) |
50 per second |
Position |
1 per second |
Battery power |
1 per second |
Differential Diagnostics (MS310) |
Thoracic respiratory motions (thorax) |
25 per second |
Abdominal respiratory motions (abdomen) |
25 per second |
Snoring sounds (snoring) |
25 per second |
Options |
Mask pressure (e.g. CPAP) |
|
Top
Parameter Specification of the Basic Diagnostics (MS300/310)
Signal/ Parameter |
Object |
Value |
Accuracy/ Comments |
SpO2 |
Range of display |
45...100 % functional SpO2 |
Increment: 1 % |
Accuracy |
2 % |
70 % < SpO2 < 100 % (< 70% unspecified) |
Internal sampling |
300 Hz |
Frequency accuracy < 100 ppm |
Average |
Adaptive |
Depending on pulse amplitude (Flexible Artefact Leveling) |
Digital filter |
50, 60 Hz and 100 Hz |
100 Hz due to neon tube |
Measuring dynamic |
Response time: 2 sec. |
Change from 84 % to 96 %, sensitivity (normal = default), PR = 75 |
|
Reach of final value: 8 sec. |
Change from 84 % to 96 %, sensitivity (normal = default), PR = 75 |
Ambient light suppression |
Up to the 14-fold of the measuring signal |
SpO2 Sensor |
Wavelength |
660 / 905 nm |
|
Max. thermal rating |
20 mW max. |
Max. rise in temperature at the application place: 2 °C |
Pulse Rate |
Range of display |
30 - 200 beats/min; |
Increment: 1 beat/min |
Accuracy |
Up to 2 % from the displayed value |
|
Average |
Adaptive |
Flexible Artefact Leveling (FAL) |
Measuring dynamic |
Reaction after max. 8 pulse beats |
Sudden change from 40 to 80 beats/min. and vice versa,
SpO2 = 96 % |
| |
Time from reaction to reach of final value: 6 sec. |
Sudden change from 40 to 80 beats/min. and vice versa,
SpO2 = 96 % |
Pulse Wave (Plethysmogram) |
Range of values |
0 – 100 % |
Resolution |
8-Bit |
Pulsation Strength (Total Pulse Modulation) |
Measurement range |
0 – 25 % |
Resolution |
8-Bit |
Respiratory Signal (Thermistor) |
Range of values |
0...16384 (-512% ... +512%) |
Resolution |
14-Bit (internal: 16-Bit) |
Filter |
0,1 Hz highpass |
Position |
Determination of 5 body positions: |
Supine |
Right |
Left |
Upright |
Prone |
Voltage of Batteries/ Rechargeable Batteries |
Resolution |
8-Bit |
Seitenanfang
Parameter Specification of the Differential Diagnostics (MS310)
Thorax |
Measurement range |
-100 .... 100% (0 ... 3,0V) |
Resolution |
10 Bit |
Transfer rate |
25 Samples/s |
Frequency range |
0,05 Hz ... 16 Hz |
Abdomen |
Measurement range |
-100 .... 100% (0 ... 3,0V) |
Resolution |
10 Bit |
Transfer rate |
25 Samples/s |
Frequency range |
0,05 Hz ... 16 Hz |
Snoring Sounds |
Measurement range |
40 ... 92 dB @ 1 mPa, 1 kHz |
Resolution |
12 Bit |
Frequency range |
50 Hz ... 1,6 kHz |
Sampling rate |
19,2 kHz |
Transfer rate
(storing the amplitude) |
25 Samples/s |
Top
MS300/310
Recording Start and Stop |
Free to program for several days |
Recording Period |
Up to 8 recordings possible |
|
|
Equipment |
Recording Period for Basic Configuration |
Memory in the Recorder |
64 MB (flash memory) |
7 days à 8 h (time left: 26:53) |
128 MB (flash memory)
Default |
7 days à 8 h (time left: 111:09) |
Optional more |
|
|
Display |
Graphical LCD display (96 x 32 Pixel) |
Data Transfer |
USB (Universal Serial Bus) Standard 1.1 |
Transfer Rate |
2,5 Mbit/s @12 Mbit/s USB clock; Standard CF, Pentium IV 1,7 GHz |
Transfer Time |
40 seconds @ 8 h recording with all sensors in the basic configuration |
Power Supply |
2 batteries 1.5-V Type AA/LR6 (Mignon)
Or alternatively
2 rechargeable batteries 1.2-V Type AA/LR6 (Mignon) with min. 1300 mAh |
|
|
Type |
Max. |
Conditions |
Charging Rate at USB |
55 mA |
75 mA |
Effective @5 V, mode: Programming and data transfer |
Charging Rate at Batteries/Rechargeable Batteries |
110 mA |
|
Effective @2,5 V
Programming mode with sensors of the basic configuration, 25 °C |
Required Charge |
900 mAh |
1000 mAh |
8 h record time, MS300 (basic configuration),
25 °C |
|
Power supply |
2 batteries |
1.5-V Type AA/LR6 (Mignon) |
Or alternatively |
2 nickel metalhydride batteries (NiMH) |
1.2-V Type AA/LR6 (Mignon) with min. 1300 mAh |
Connections |
SpO2 Sensor: |
Six-pin M&S SIM flat connection |
Airflow Sensor: |
Six-pin M&S SIM flat connection |
Options (AUX): |
Six-pin M&S SIM flat connection |
|
Interface |
Jack at the MS300/310 |
Mini USB jack ("Receptable") Type B |
Enclosed connection cable to a standard PC |
USB connector ("Plug") Type A at USB connector ("Plug") Type B (screened high speed cable) |
Dimensions (L x W x H) |
11,0 cm x 6,2 cm x 2,8 cm |
|
Weight |
36 grams |
MS300 recorder plate (without plate electronics) |
120 grams |
MS300 including recorder plate, without batteries |
Top
Environmental Conditions for Transport and Storing (MS300/310)
|
Min. |
Max. |
Unit |
Temperature |
-20 |
+50 |
°C |
Humidity |
30 |
80 (noncondensing) |
%RH |
Air Pressure |
700 |
1060 |
hPa |
Top
Operating Conditions (MS300/310)
|
Min. |
Max. |
Unit |
Temperature |
+10 |
+40 |
°C |
Humidity |
30 |
80 (noncondensing) |
%RH |
Air Pressure |
700 |
1060 |
hPa |
Top
|