Müller & Sebastiani Elektronik GmbH
About Us Test & Measuring Devices Medical Devices Online Shop Services Contact Us

Deutsche Version | Home | Sitemap | Imprint



Zeus Defibrillation Simulator
phantom 320 ECG Simulator
MS410 ECG Simulator
MS400 Calibration Signal Generator
Zeus Defibrillation Simulator

 


  Services -> Frequently Asked Questions

Frequently Asked Questions

Zeus Defibrillation Simulator

MS410 ECG Simulator

phantom 320 ECG Simulator

General


Zeus Defibrillation Simulator

Q: Is there a possibility for the functional testing of the device?

A: You can test the output voltage with a high voltage probe and an oscilloscope. WARNING! Please note the safety regulations for measurements at high voltage! The discharge curve of the capacitor at the oscilloscope can be compared to the values in the calibration report.

Q: When does the device have to be sent back for calibration?

A: We recommend a calibration after approximately 5000 discharges or after 24 months.

Q: What is the maximum energy of the device at the testing output?

A: The maximum energy is calculated from the equitation:
W = ½ * C * U² = ½ * 32 uF * 5000² = 400 J

Which model variants comply with which standards?

A: Version I (500 uH/100R) part number 59.101 complies with the standards
DIN EN / ISO 60601-1: 03-1996 / 17h
DIN EN / ISO 60601-2-25: 04-2001 / 17.101, 51.102, 51.103

Version II (25mH/ 400R) part number 59.102 complies with the standards
DIN EN / ISO 60601-2-27: 03-2004 / 17.101 aa
DIN EN / ISO 60601-2-26: 01-2004
DIN EN / ISO 60601-2-49: 12-2002

Q: Is there a version for 120 V supply voltage?

A: No, you need a voltage converter for other supply voltages.

Q: Can the device also be calibrated in the USA or in other countries?

A: No, the device must be sent back to us for calibration.

Q: For a measurement with the oscilloscope at the measurement output of the Zeus voltages up to 8 Vss occur although no disruptive discharge can be noticed.

: The time basis of the oscilloscope is not correct (under 1 ms). EN / ISO 60601-1: 03-1996 for testing in accordance with point 17h shows in figure 50 the RC element for the recording of the discharges. R2 = 100 k? and C1 respectively C2 = 1 uF results in a time constant of 100 ms!
All signals which are recorded with a lower time basis than mentioned above are to be considered as pure high voltage interfering signals.

Q: How can I use the "Measuring-In"-Input?

A: This input serves for the measurement of possible voltage at tangible enclosure parts of the device under test. See figure 50 in EN / ISO 60601-1: 03-1996

Q: Does the Zeus provide patient simulation?

A: The patient is simulated as R1 and R3 as described in the corresponding standards. See the block diagram in the Zeus instruction manual and the corresponding figures in the standards.

Top   


MS410 ECG Simulator

Q: In which languages is the MS410 available?

A: In the "Adjustment" menu you can select "German" or "English".

Q: Does the MS410 include all testing ECGs of the standard DIN EN / ISO 60601-2-51?

A: No, the MS410 includes the signals of the so-called CTS-Atlas (appendix HH, DIN EN / ISO 60601-2-51: 02-2004) as well as artificial but adapted to human ECG signal forms.

Q: When does the device have to be sent back for calibration?

A: We recommend a calibration after 24 months.

Q: In the MS410 instruction manual, page 22 (26-27) the 24 types of ECG signals are described. Can these signals be used for testing in accordance with the standard IEC 60601-2-51 point 50.102.3?

A: These ECG signals are generated artificially in spite of their "natural" appearance. They have been generated to enable a quick and simple testing of an ECG device. Regrettably they are not suitable for a testing in accordance with IEC 60601-2-51: 02-2004, point 50.102.3. For these tests the signal curves need to be used as described in the standards.

Q: Are the amplitudes of the MS410 continuously adjustable?

A: No, the amplitudes are set fixed in the particular programs.

Q: The display is dark and difficult to read.

A: Hold the MS410 away from the body and watch the display slanted from below.
You can dimly read the menue and change the settings:

  • Select Adjustment from the mainmenue (mainmenue arrow buttons).
  • Press the submenue button until Contrast is displayed (submenue arrow button, 2 x).
  • Press the up-arrow button until the contrast is optimal.

Top   


phantom 320 ECG Simulator

Q: Is the phantom 320 also available in an US version with the corresponding AHA colours?

There are 2 versions available as described in the standards DIN EN / ISO 60601-2-51: 02-2004 point 101 und DIN EN / ISO 60601-2-27: 03-2004 point 1:
Version I with Code 1 (Europe) – part number 59.009
Version II with Code 2 (America) – part number 59.010

Q: When does the device have to be sent back for calibration?

A: We recommend a calibration after 60 months.

Top   


General

Q: Are there distributors of the products abroad?

A: No, please send your order directly to us.

Q: Is the sale and dispatch carried out also to the USA or only to Europe?

A: We sell and send to the whole world.

Q: Can you pay by credit card?

A: Regrettably no. The payment is only possible using bank transfer or by check.

Q: Who can carry out the calibration?

A: Please send the device to us. There are no companies which are authorized by us for the calibration.

Q: Are the devices delivered with the calibration certificate?

A: The MS400, MS410 and the Zeus are always delivered with the calibration certificate (included in the price). For the phantom 320 the certificate needs to be ordered and paid separately.

Q: How much time takes the calibration?

A: We need approximately 3 to 4 days for the calibration of the device.

Top   

 

© 2004 - 2012 Müller & Sebastiani Elektronik GmbH - Webmaster